Title:


CONDOR – a heterodyne receiver at 1.25-1.5THz

Abstract:
(Your abstract must use Normal style and must fit in this space)

The CO N+ Deuterium Observations Receiver (CONDOR) is a heterodyne receiver that operates between 1250 – 1530 GHz. Its primary goal is to observe star-forming regions in CO, N+, and H2D+ emission.

The instrument follows the standard heterodyne design. It uses a solid state local oscillator (LO), whose signal is overlaid with that of the sky using a Martin-Puplett interferometer. The heart of the receiver is a superconducting NbTiN hot electron bolometer (HEB) (Munoz et al. 2006). The bolometer has an area of 0.25 x 2.8 microns and is mounted on a SiN membrane in a waveguide mixer block. To facilitate operation at remote sites, CONDOR is the first receiver that cools the HEB with a closed-cycle system. Since HEBs are particularly sensitive to temperature fluctuations as well as modulations in LO power, we use a Pulse Tube Cooler, which has less vibration than, e.g., a Gifford McMahon cooler. In order to further minimize vibrations and temperature fluctuations, the mixer and first amplifier are mounted on a separate plate connected via flexible heat straps to the 4K stage. CONDOR has an intermediate frequency (IF) of about 1.0 -1.8 GHz. We consistently obtain receiver noise temperatures below 1800 K and minima in the spectral Allan variances at 25 - 35~s, which is approximately the optimum individual on-source integration time.

In November 2005, CONDOR was successfully commissioned on the 12-m Atacama Pathfinder Experiment (APEX) telescope and the first astronomical observations were performed (see contribution “CONDOR Observations of High Mass Star Formation in Orion” by Volgenau).


see Wiedner et al. (2006).)\\


\

located at an elevation of 5100~m in the Atacama desert in Chile.

Pointing observations were preformed on the Moon and Mars.

The first spectral line observations were obtained of CO J=13-12 emission

at 1497~GHz from several sources in Orion.

(For details on the first light observations

see Wiedner et al. (2006).)\\

integration time (see Schieder \& Kramer 2001).

As expected, the Allan variance is

dominated by the mixer and the LO and not the IF.

(The minimum time of the total power Allan variance is about

10s over a 1~MHz bandwidth. This variance

is important for continuum observations.)


We consistently obtain receiver noise temperatures below 1800~K

using hot (ambient temperature) and cold (liquid nitrogen) calibration

loads. However, reliable noise temperatures as low as 1400~K

were seen on dry days and with optimal tuning.

Receiver temperatures are flat over the entire 800~MHz band.

We measured spectral Allan variances by computing the variance

of the calibrated differences of consecutive spectra. These variances

are important for spectral line observations because they indicate

baseline ripples caused by temporal deformations of the passband.

We obtain minimum Allan variances at 25 - 35~s,

which is approximately the optimum individual on-source

integration time (see Schieder \& Kramer 2001).

As expected, the Allan variance is

dominated by the mixer and the LO and not the IF.

(The minimum time of the total power Allan variance is about

10s over a 1~MHz bandwidth. This variance

is important for continuum observations.)

In November 2005, CONDOR was successfully commissioned on the 12-m

Atacama Pathfinder Experiment (APEX) telescope, which is

located at an elevation of 5100~m in the Atacama desert in Chile.

Pointing observations were preformed on the Moon and Mars.

The first spectral line observations were obtained of CO J=13-12 emission

at 1497~GHz from several sources in Orion.

(For details on the first light observations

see Wiedner et al. (2006).)\\


\smallskip

\noindent

References:\\

Schieder, R. \& Kramer, C. 2001, 373, 746\\

Wiedner, M. C., et al. 2006, A\&A accepted\\